Download Autosampler 3950 User Manual
Transcript
24 PASA™ Loop Injection Principle maximum sample volume is at 50 % of the loop volume minus 1.5 times the needle volume. Air needle and sample needle The autosampler uses a system of two telescopic needles, one that pierces through the cap of the sample vial – the air needle – and one that extracts the sample – the sample needle. Using a syringe, the sample is aspirated through the two needles out of the sample vial while under pressure and into the sample loop. To prevent the syringe from becoming contaminated, a buffer tube is situated between the syringe and the valve. Using washing solution, sample residue is removed from the sample needle and buffer tube. Full Loop Injection Note A flushing process takes places after every injection. Functional schematic Explanations 1. Initial conditions: The valve switches to the INJECT position. Sample needle and air needle are inserted into the vial. The air needle creates pressure, which prevents air and steam bubbles from developing. Fig. 10 Full loop injection: Initial conditions KNAUER Autosampler 3950 700 bar, user manual V1511A, version 1.0