Download PROOFMASTER S

Transcript
Lindholts Fourniturehandel ApS. | Sportsvej 6 | 8930 Randers NØ
Note:
As explained above, the leak test can be performed with PROOFMASTER S according to the ISO
standard with results being in good agreement. However, since the test principle of
PROOFMASTER S does not correspond to the ISO 2281 standard, it is not possible to ensure a 100%
conformity.
6.7 MEASUREMENT RESOLUTION (PRECISION)
The measurement resolution determines the smallest measurable unit for deformation reversal.
With PROOFMASTER S, the measurement resolution is determined by the programmed waterresistance limit and the selected test mode (Lab, Stnd, Prod, see Chapter 4.7 Tightness limits, page
19). With the pre-set programs, the instrument measures with a 25% resolution of the waterresistance limit, i.e. 4 times more accurately than the programmed tolerance limit.
The higher the measurement resolution, the more accurate the measured leak rate with
correspondingly longer measuring times.
The value for customer-specific programs can be selected between 5% (Labo), 25% (Stnd) and 40%
(Prod) measurement resolution values.
Tlf. 86 42 52 99 | Fax 89 41 37 45 | lindholts@email.dk | www.lindholts.dk