Section 8: Calibration
Section Overview
Introduction
This section describes the step-by-step procedure for calibration of the
Quintus. The instrument has been calibrated by Boule prior to shipment.
Good laboratory practice, however, requires regular checks and calibration of
the measured parameters. It is recommended to calibrate the instrument when
needed.
Contents
This section contains the following topics:
Topic
Preparations before calibration
Calibration
See Page
65
66
8.1 Preparations before calibration
Before
Calibration
Important
Warning
· It is recommended that the performance of the Quintus system is checked
daily with certified blood controls authorized by Boule.
· Analyze control blood once in the open tube mode and compare results with
the assigned values prior to calibration.
· Verify that nothing is erratic with the control blood, the reagents, or the
instrument before calibrating instrument.
· Verify that instrument maintenance/cleaning is current. (See Sections 9.1 –
9.3)
· Prior to calibration print Calibration Log. Select [CALIBRATION/QC] from
Main menu, then [CALIBRATION], then [VIEW CALIBRATIONS], select
latest calibration, and then [PRINT/SEND].
· The user should be thoroughly familiar with the analyzer system and the
calibration procedure before performing calibration.
· Refer to the Calibrator Product Insert for complete instructions for handling
and use of blood calibration materials.
· Never use an open vial longer than recommended by the manufacturer or
subject any vial to excessive heat or agitation.
· As there are no assurances of the absence of HIV, Hepatitis B or C viruses
or other infectious agents in blood samples, controls, and calibrators these
products should be handled as potentially biohazardous.
· Always wear protective gloves and goggles. Follow local regulations.
65