bio bas. 1- User’s manual
1.INTRODUCTION
1.1 GENERAL:
SPINREACT, bio bas.1 coagulometer is an instrument created to determinate the most important
parameters run in plasma coagulation methods:
Prothrombin Time or Quick Time (PT), Activated Partial Thromboplastin Time (APTT), Thrombin
Time (TT), Fibrinogen Concentration (FB), and all coagulation methods whose principle is the
determination of a clotting time.
bio bas.1 stores the calibration curves for Prothrombin Time and Fibrinogen Concentration, which
means no calculation is needed to be done by the user because it gives the result in time, activity,
concentration, RATIO and INR. (different parameters, depending on the method).
1.2 THEORY OF THE INSTRUMENT FUNCTIONING:
bio bas.1 has an optical measurement system which detects a sudden variation of the optical
density when a clot is formed.
The chronometer and the stirring system are activated by a sudden change of the optical density.
This permits the initiation of the time measurement when the starter is added and to stop the
measurement time at the moment the clot is formed.
The continuous mixing guarantees a perfect homogenisation and makes possible the measurement of
low concentrations of Fibrinogen by grouping the fibrin filaments in the centre of the optical path.
The system has a programmable security time during which variations of the optical density, when
the reagent and the plasma are still in the homogenisation phase, can not activate the detection cell.
1.3 GENERAL DESCRIPTION OF THE INSTRUMENT:
Dimensions: 16.5 x 29 x 12.8 cm.
Voltage and frequency: 220 V
50 Hz, 60 Hz
Power data: 100 VA
bio bas.1 has the following elements:
Incubation zone for 10 cuvettes and one reagent.
Reading zone.
Keyboard.