Download Manual C.diff-Type AS-1 Kit
Transcript
www.alere-technologies.com www.alere.com Discard collection tube with liquid. Place the spin column in a new 2 ml collection tube (provided with the kit). Add 500 µl Buffer AW1. Centrifuge (8,000 rpm, 1 min) at room temperature. Discard collection tube with liquid. Place the spin column in a new 2 ml collection tube (provided with the kit). Add 500 µl Buffer AW2. Centrifuge (14,000 rpm, 3 min) at room temperature. The membrane of the spin column should be dry, and all liquid should be in the collection tube. Discard collection tube with liquids. Place the spin column in a clean 1.5 ml tube (not provided with the kit). Add 100 µl Buffer AE (or PCR grade distilled water) directly onto the membrane of the spin column. Incubate at room temperature for 1 min to elute DNA. Centrifuge (8000 rpm, 1 min) at room temperature. Optional: Add another 100 µl Buffer AE (or PCR grade distilled water) directly onto the membrane, incubate at room temperature for 1 min and centrifuge again. Discard the spin column. Please note: Ethanol from Washing Buffers strongly inhibits the enzymes used in the assay. Contamination with Washing Buffer might occur during the elution of prepared DNA by drops adhering to the spin columns funnels. Therefore these funnels should be gently touched and dried with sterile filter paper or wipes prior to the elution step. Alternatively, prepared DNA can be heated shortly to evaporate ethanol (e.g. 10 min at 70 °C). C.diff-Type AS-1 Kit 05_16_04_0019_V01_Manual C.diff-Type AS-1 Kit 10