Variable and Calibration setting table FUNCTION AdR SER bAU PAR SR2 OUt RtC FIL SPd FRZ SAF bRt PRN DESCRIPTION Set unit address (0-255) Serial Com:1 Baudrate (1200,2400,4800,9600,19200) Parity ( 0=8,None,1 1=7,Even,1 ) Serial Com:2 (1=1200,2=2400,3=4800,4=9600) Output feature (N/A) Real Time Clock (N/A) Weight Filter (0-4) ADC read Speed (0-4) Weight Freezing (N/A) Power Safe Mode (0=Disable, 1=Enable) Display Brightness (0-4) Print Interlock Weight DEFAULT 0 3 1200 1 1 0 0 1 4 0 0 3 10 Valid when Short CAL.Jumper dEC CAP StP AZR ZER POZ StI GAN bt1 bt2 bt3 bt4 bt5 bt6 RFP CAL-0 CAL-1 CAL-2 CAL-3 AdC-dISP SAVE Decimal point Capacity Division Step (1,2,5,10,20) Auto Zero Maintennance % Zero limit ( 0,1,2,5,10 ) Power On Zero (0=Disable, 1=Enable) Still Weight or Statble level ADC Gain (0-4) Lock button No.1 (FUNC) 0=Unlock,1=Lock Lock button No.2 (MODE) 0=Unlock,1=Lock Lock button No.3 (CLEAR) 0=Unlock,1=Lock Lock button No.4 (TARE) 0=Unlock,1=Lock Lock button No.5 (ZERO) 0=Unlock,1=Lock Lock button No.6 (PRINT) 0=Unlock,1=Lock RF Protect feature (Option) Calibrate Zero Calibrate Span Calibrate Zero and Span Calibrate Manual Setting Display ADC value Save setting (Press [ENTER] to save) 0 1000 1 1 10 1 3 1 0 0 0 0 0 0 0 -