Measurement modes
HR, BP, SpO2 panel
The basic cardiologic parameters are monitored during the whole test
Main window
The most important metabolic and breathing parameters are monitored:
VO2, VCO2, VE
Spirogram
The spirogram shows the breathing, tidal volume, breath rate, and the
shifting of the breathing median
Timer
The upper counter shows the total duration of the test and the lower
counter shows the duration of the steady state
Parameter table
Values of the vital parameters are shown during the whole test.
Preparation
It is inevitable to inform the patient as described in the in the chapter
Preparations.
The following steps should be done:
• Lay the patient down to the examination bed
• The patient has to find a position whish is comfortable enough to
lay without any movement for an hour
• While the patient is laying down make the calibration of the system
• Put the facial mask on the patient
Measurement process
• Select under the [Protocol] tab in the [Ergometer type] section the
[Tidal test] option
• To start the test push the [Monitor] button
• The patient has to refrain from any movement and from speaking
as well. Usually 20 ÷ 30 minutes are requested to reach the stable
metabolic level.
• Push the [Start] button as the VO2, VCO2, VE, RER parameters
are getting stable
• The system indicates the stable intervals by frames. The REE value
is updated continuously. If all the conditions are met in a 5 minute
interval the system indicates it:
Steady state criteria met.
• Push the [Done] button
• Results are shown at the [Result] tab
ES-39