Download - Murdoch Research Repository
Transcript
(Shimaden). The slow fall function slows the fall of the output in the event of sudden change to the input signal, similar to the slow rise function. Functions available in the new Thyristors, it may not be desirable to use them as it may cause confusion amongst experimental users, unless they are well aware of these functions. 4.2.2 New 24VDC Power Supplies New 24VDC power supplies were purchased as the previous 24VDC power supplies were no longer manufactured. Unfortunately there was no manual available on the old 24VDC power supplies. Based on the label on the old power supply, its function was 240VAC input and 5Amps 24VDC output, as shown in figure 34. The new 24VDC power supplies (Mean Well, SP-150-24) have an 85~264VAC input and 6.3A 24VDC output, as shown in figure 33. In addition the new power supply features protections for short circuit, overload, over voltage and over temperature. Figure 31-New 24VDC power supply Figure 32-Old 24VDC power supply 4.2.3 Transformer 5 Primary Circuit Breaker Current Rating The circuit breaker used on the primary side of transformer 5 on the old power board uses a 16A current rating. However a 10A current rating was documented for the transformer, as shown in Figure 35. The 16A current rating circuit breaker might be due to the high inrush current. However it was decided that a 10A current circuit breaker would be placed in the new power board. This would allow future users to see if the circuit breaker trips when the power to the power board is switched on, showing that the inrush current is 33