Model 421 Rear Panel
Œ
1 - Line Input Assembly

2 - Serial I/O Interface
3 - Corrected Analog Output

4 - Monitor Analog Output
5 - Probe Input
6 - Relay Terminals
‘
Ž

Hall Probes*
L
2.5"
A
Axial Probes
+B
fi
D
0.36" diameter
±0.030"
Cable length
6.6 feet
Model
L
D
A
MMA-2502-VH
2" ±0.063"
0.25" dia
±0.006"
0.010" ±0.005"
4" ±0.125"
0.187" dia
±0.005"
MMA-2502-VG
2" ±0.063"
0.25" dia
±0.006"
0.010" ±0.005"
Aluminum
MNA-1904-VG
4" ±0.125"
0.187" dia
±0.005"
0.005" ±0.003"
Fiberglass
epoxy
MNA-1904-VH
Active area
Stem material
Frequency
range
Aluminum
0.030" dia
(approx)
0.005" ±0.003"
Fiberglass
epoxy
Usable full
scale ranges
Corrected
accuracy
(% of reading)
30 G, 300 G,
3 kG, 30 kG
±0.25% to
3 0 kG
300 G, 3 kG,
3 0 kG
±0.15% to
3 0 kG
DC and
10 Hz to 400 Hz
Operating
temperature
range
Temperature
coefficient
(maximum)
Zero
Temperature
coefficient
(maximum)
Calibration
±0.09 G/°C
-0.04%/°C
±0.13 G/°C
±0.005%/°C
0 °C to +75 °C
+B
L
2.5"
Transverse Probes
A
W
fi
T
0.36" diameter
±0.030"
Cable length
6.6 feet
Model
L
T
W
MMT-6J04-VH
4" ±0.125"
0.061" max
0.180" ±0.005"
MNT-4E04-VH
MMT-6J04-VG
MNT-4E04-VG
4" ±0.125"
4" ±0.125"
4" ±0.125"
0.045" max
0.061" max
0.045" max
A
0.150" ±0.005" 0.150" ±0.050"
Frequency
range
Usable full
scale ranges
Corrected
accuracy
(% of reading)
Aluminum
DC
30 G, 300 G,
3 kG, 30 kG
±0.25% to
3 0 kG
Rigid glass
epoxy
DC and
10 Hz to 400 Hz
300 G, 3 kG,
3 0 kG
±0.15% to
3 0 kG
Active area
Stem material
0.040" dia
(approx)
0.180" ±0.005"
Aluminum
DC
0.150" ±0.005"
Rigid glass
epoxy
DC and
10 Hz to 400 Hz
Operating
temperature
range
Temperature
coefficient
(maximum)
Zero
Temperature
coefficient
(maximum)
Calibration
±0.09 G/°C
-0.04%/°C
±0.13 G/°C
±0.005%/°C
0 °C to +75 °C
+B
L
S
A
2.5"
W
Flexible Transverse Probes
fi
Cable length
6.6 feet
"L" = 3", "S" = 0.375"
0.36" diameter
±0.030"
0.125" diameter
max
T
Model
W
T
A
Active area
Stem material
Frequency
range
Usable full
scale ranges
Corrected
accuracy
(% of reading)
Operating
temperature
range
Temperature
coefficient
(maximum)
Zero
Temperature
coefficient
(maximum)
Calibration
MFT-3E03-VH
0.135" max
0.025" max
0.125" ±0.006"
0.040" dia
(approx)
Flexible tubing
DC and
10 Hz to 400 Hz
30 G, 300 G,
3 kG, 30 kG
±0.25% to
3 0 kG
0 °C to +75 °C
±0.09 G/°C
-0.04%/°C
300 G, 3 kG,
3 0 kG
±0.15% to
3 0 kG
±0.13 G/°C
-0.005%/°C
MFT-3E03-VG