EMISSION TEST
4.1 CONDUCTED EMISSION MEASUREMENT
4.1.1 LIMITS OF CONDUCTED EMISSION MEASUREMENT
TEST STANDARD: EN 55022
FREQUENCY (MHz)
0.15 - 0.5
0.50 - 5.0
5.0 - 30.0
NOTE:
Class A (dBuV)
Quasi-peak Average
79
66
73
60
73
60
Class B (dBuV)
Quasi-peak Average
66 - 56
56 - 46
56
46
60
50
(1) The lower limit shall apply at the transition frequencies.
(2) The limit decreases in line with the logarithm of the frequency in the range of 0.15
to 0.50 MHz.
(3) All emanations from a class A/B digital device or system, including any network of
conductors and apparatus connected thereto, shall not exceed the level of field
strengths specified above.
4.1.2 TEST INSTRUMENTS
DESCRIPTION &
MANUFACTURER
MODEL NO.
SERIAL NO.
CALIBRATED
UNTIL
ROHDE & SCHWARZ
Test Receiver
ESCS 30
100287
Feb. 10, 2007
Line-Impedance Stabilization
Network(for EUT)
ESH3-Z5
848773/004
Oct. 24, 2006
Line-Impedance Stabilization
Network(for Peripheral)
ENV-216
100072
Oct. 24, 2006
RF Cable (JETBAO)
RG233/U
Cable_CA_01
Jul. 19, 2007
50
1
Oct. 08, 2007
ADT_Cond_V7.3.2
NA
NA
Terminator
Software
NOTE: 1. The calibration interval of the above test instruments is 12 months and the calibrations
are traceable to NML/ROC and NIST/USA.
2. The test was performed in ADT Shielded Room No. A.
3. The VCCI Con A Registration No. is C-817.
Report No.: CE951004H04F
Reference No.: 960130H03, 981222A14
14
Report Format Version 3.0.1