CircuLex Mouse FABP3/H-FABP ELISA Kit User’s Manual For Research Use Only, Not for use in diagnostic procedures On ly! TM Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in four separate assays to assess inter-assay precision. • Inter-assay (Run-to-Run, n=4) CV=9.2, 4.4, 4.8% Mouse FABP 3 conc. (ng/ml) 1 2 3 4 MAX. MIN. MEAN S.D. C.V. Serum 2 57.3 59.0 58.7 53.5 59.0 53.5 57.1 2.53 4.43% Serum 3 11.5 12.6 12.8 12.5 12.8 11.5 12.3 0.59 4.80% rp os e Serum 1 92.3 102.0 115.5 105.4 115.5 92.3 103.8 9.55 9.20% 3. Linearity Pu To assess the linearity of the assay, samples containing and/or spiked with high concentrations of mouse FABP3/H-FABP were serially diluted with the Dilution Buffer to produce samples with values within the dynamic range of the assay. Linearity ce ● Serum 1 ▲ Serum 2 ◆ Serum 3 1.5 1.0 en 2.0 er Mouse FABP 3 conc. (ng/ml) 2.5 ef 0.5 0.0 rR 0 Fo Cat#: CY-8055 0.5 1 1.5 2 2.5 Serum Dilution Ratio (1/100) 13 Version#: 120710