Routine Maintenance
Set the Temperature for the Stack Holder
The temperature setting for the stack holder is shown on the front of the stack holder power
supply. Change the temperature if the default is not applicable.
Figure 4-17 Stack Holder Power Supply Control Panel
1. Press P for less than two seconds.
“SP1” is displayed.
2. Press the Up and Down arrows to reach the desired temperature.
3. Press P again to set the temperature.
Replace the Autosampler Fuse
Required Materials
• 6.3 A 250 VAC 5 mm x 20 mm fuse
1. If the system is not already off, turn off the power switch on the autosampler power
supply.
2. Remove the fuse from the back of the Z arm of the autosampler (Figure 4-18).
Figure 4-18 HTC-xt PAL Autosampler Z-Arm—Rear View
1
Item Description
1
Fuse
3. Replace the fuse.
4. Power up the system.