Troubleshooting Table 5-2 Flow Control System (Continued) Symptom Possible Cause System pressure (Pc) is Incorrect k-values low and flow rate is OK but pump pressures (Pa and Pb) are high Flow module is plugged Corrective Action Calibrate the flowmeters (refer to Calibrate Flowmeters on page 54). Contact AB SCIEX Technical Support for assistance. System pressure (Pc) is very high Tubing or fitting is plugged Starting at the waste tanks, remove each tube or fitting, one at a time, until the pressure drops dramatically. Replace the plugged item. Excess flow noise Trapped air in pump Purge the pumps (refer to step 3 in Purge Mobile Phases on page 51). Pump controller out of tune Contact AB SCIEX Technical Support for assistance. Pump controller out of tune Contact AB SCIEX Technical Support for assistance. Pump time response is set incorrectly Adjust the pump time response in the Flow Metering and Control section of the Hardware Diagnostics dialog. Measured flow does not follow the flow profile Pump pressures (Pa and Incorrect gain setting for Pb) maximized to pressure <12 000 psi at 100% pump power Verify the pump pressures are 2800 psi/V in the Calibration Values tab of the Hardware Diagnostics dialog. Incorrect zero setting for pressure sensors Re-initialize the pressure transducers (refer to Re-initialize the Pressure Transducers on page 52). In-line air pressure too low. Verify that the air supply has an air pressure of 100 psi. Symptom Possible Cause Corrective Action Column responds very slowly when changing temperature Oven malfunction Contact AB SCIEX Technical Support for assistance. Temperature reads 47 Oven unplugged Verify the column oven is plugged in. Table 5-3 Column Oven Operator Guide RUO-IDV-05-1129-B | D5033433 G Eksigent MicroLC 200 Plus System 99 of 140