Chapter 5. One-Color Array Scanning
Chapter 5 describes the protocol for scanning one-color NimbleGen arrays with
a GenePix 4000B Scanner and associated software. Keep arrays in a dark
desiccator until you are ready to scan them.
1. Turn on the scanner using the power switch on the back right side.
2. Launch the GenePix software 10 minutes before scanning to allow lasers to
warm.
3. Open the scanner door and the slide carriage.
4. Using a NimbleGen Compressed Gas Nozzle, gently blow compressed inert
gas across the slide to remove any dust or debris from the array. Do not use
canned aerosol compressed air for this purpose.
5. Place the slide in the slide carriage so that the array is face down and the
barcode end is closest to you.
6. Move the black lever on the left side of the
carriage to the left until the slide is lying flat in
the carriage.
7. Release the lever so that the slide is gently pushed
to the right side of the carriage and held firmly.
8. Close the carriage (you should hear a click) and
slide the scanner door shut.
9. Open the Hardware Settings dialog box (right
side, second icon from the bottom or Alt + Q).
Select the following settings for scanning:
•
•
•
•
•
•
Wavelength = 532 (Cy3)
PMT Gain = 500
Power (%) = 100
Pixel size (μm) = 5
Lines to average = 1
Focus position (μm) = 0
Figure 11: Hardware
Settings Dialog Box
10. Under Image on the left side of the screen, select the 532 wavelength.
11. Preview the entire slide by clicking the preview scan icon (or Alt + P). This
will allow you to locate the array(s) on the slide.