Download CPP 5Gs User`s Manual
Transcript
80-column result printout 1 - 80-column result printout 30/09/2005 11:31 ISL CPP 5Gs ALAN ID:ISL CPP 5Gs Cooling unit S/N:M100B3#323/7 Sample ID:12181 Program : 7 D97 Pour: -27.0°C Date: 30/09/2005 No: Software V 2.1/ 2.1 S/N: 223 Exp. pt. : -24.0°C Operator : AM Time: 11:01 Program No: 7 Name: D97 Type of analysis: Pour First test at EP+ 9°C Stop 3 tests after EP: NO Alarm at -150.0°C Test frequency: 3.0 Detection level: 350 Stop at -150.0°C Cooling profile No: 7 Name:D97 Step : 1 Sample preheating at 45.0°C for 0 min 2 If EP > -33.0°C then step 3 else step 4 3 Jacket level at 24.0°C until sample < 27.0°C 4 Jacket level at 6.0°C until sample < 15.0°C 5 Jacket level at 0.0°C until sample < 9.0°C 6 Jacket level at -18.0°C until sample < -6.0°C 7 Jacket level at -33.0°C until sample < -24.0°C 8 Jacket level at -51.0°C until sample < -42.0°C 9 Jacket level at -69.0°C until sample < -60.0°C 10 Jacket level at -87.0°C until sample < -78.0°C 11 Jacket level at -105.0°C until sample < -96.0°C 12 Jacket level at -123.0°C until sample < -114.0°C 13 Final heating, jacket temperature 25.0°C DOCV223A001-F CPP 5Gs User's Manual Page 75