Download Agilent Technologies 6111A User`s guide
Transcript
Absolute Measurement Examples 6 RF Synthesizer Using EFC This measurement example will help you measure the absolute phase noise of an RF synthesizer using EFC. C AU T I O N To prevent damage to the test set’s components, the input signal do not apply the signal input connector until the input attenuator has been correctly set for the desired configuration, as shown in Table 31. Apply the input signal when the connection diagram appears Required equipment This measurement requires an Agilent 8257x with an EFC Input port, in addition to the phase noise test system and your DUT. (For more information, refer to the section “Selecting a reference source" on page 132.) You also need the coaxial cables and adapters necessary to connect the DUT and reference source to the test set. NOTE To ensure accurate measurements allow the DUT and measurement equipment to warm up at least 30 minutes before making the noise measurement Defining the measurement 1 From the File menu, choose Open. 2 If necessary, choose the drive or directory where the file you want is stored. 3 In the File Name box, choose “RFSynth_EFC.pnm”. See Figure 120. Figure 120 Select the parameters definition file 4 Click the Open button. 174 Agilent E5505A User’s Guide
Related documents
Agilent Technologies E5503A User`s guide
American DJ DCD-PRO300 Specifications
E5500 Phase Noise Measurement System Version A.02.00
Agilent Technologies 4352B Instruction manual
Test and Measurement Instrument Security
Data Sheet - Agilent Technologies
view data sheet
Agilent Technologies E5500 Series Specifications
Agilent Technologies E5503A Installation guide
Agilent Technologies E5500 Series Installation guide
"service manual"
Service Manual