5
Performance Tests and Calibration
Step
Test Function
Cal Range
Input
Cal Item
U1241A
6
7
Turn the rotary switch to
μA position
Press
to go to
μA
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
1000 μA
1000 μA
1000.0 μA
10000 μA
10000 μA
10000 μA
1000 μA
50 μA, 70 Hz
50.0 μA
100 μA, 70 Hz
100.0 μA
1000 μA, 70 Hz
1000.0 μA
1000 μA, 70 Hz
1000 μA
10000 μA, 70 Hz
10000 μA
function
10000 μA
8
9
Turn the rotary switch to
mA position
Press
to go to
mA
oPEn
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
100 mA
100 mA
100.00 mA
1000 mA
320 mA
320.0 mA
100 mA
5 mA, 70 Hz
5.00 mA
10 mA, 70 Hz
10.00 mA
100 mA, 70 Hz
100.00 mA
100 mA, 70 Hz
100.0 mA
320 mA, 70 Hz
320.0 mA
function
1000 mA
54
U1242A
oPEn
U1241A/U1242A User’s and Service Guide