Download Agilent Technologies ESG Specifications
Transcript
ESG Family Signal Generators Performance Tests 14. Custom I/Q RF Modulation Quality (Option UN8 only) 14. Custom I/Q RF Modulation Quality (Option UN8 only) This automated test verifies the RF modulation quality of the Option UN8 signal generator’s internal real time I/Q baseband modulation. A vector signal analyzer is connected to the signal generator’s RF output. The Option UN8 internal baseband generator modulates the RF carrier in each of the TDMA digital modulation formats. The parameters of the individual TDMA formats common to Options UN3, UN4 and UN8 are tested elsewhere. The vector signal analyzer measures the appropriate error parameter for the modulation generated: • EVM for QPSK, π/4DQPSK, and QAM16 • Phase error for MSK • Frequency Shift Error (%) for FSK2 Recommended Equipment • MD/MDC-174 Mixer (0.001 to 2.8 GHz) • HP/Agilent 8663A Signal Generator • HP/Agilent 89441A Options AY9, AYA, AYH, UFG or UTH Vector Signal Analyzer Equipment Setups Figure 4-26 Custom I/Q RF Modulation Quality ≤ 2000 MHz Test Setup Calibration Guide 4-29