Commissioning 5.4.6 PR 5220 Operating Manual Calibration with Weight (by Load) Select [by load] for calibration using weight. Place CAL weight The weight value for the calibration weight must be entered in a separate on the scale window. and enter value 2000 kg After applying the weight, enter the weight value and confirm with . The weight unit for the calibration to change) may differ from the unit in the instrument; conversion is automatic. weight (press Afterward, the following message is displayed: Setting SPAN by load Weight value, weight unit and measuring signal in mV/V corresponding to this value are displayed in the [Calibrated at] line. Max Scale interval Deadload at Max at Calibrated at Sensitivity by load WP A/Calibration 3000 d 3000 d 165.11 kg 3000.00 kg 2000 kg 876.97 by mV/V by data 3000 kg 1 kg 0.057920 1.052369 0.701579 4.209600 Linear. Test The scale is not stable. Remedial action: Check the mechanical function of the scale; adapt the filter setting; reduce the resolution; if necessary, adapt the stability conditions. Set SPAN failed No stability Set SPAN failed Load below deadload The weight on the scale is less than the dead load after input of the weight value. The next step is calculation of the test value with [Test] (see Chapter 5.4.11), and calibration is completed with (see Chapter 5.4.12). Determining span without weight Max Scale interval Deadload at Max at Not calibrated Sensitivity by load 38 WP A/Calibration 3000 d 3000 d 3.00 kg 3000.00 kg by mV/V 833.33 by data 3000 kg 1 kg 0.001000 0.000000 4.000000 Linear. Test Sartorius