Installation Operating Instructions Chapter 4 ICR890 4.3 Installing and adjusting the device The position of the components on the installation frame is specified project specifically in a dimensional sheet (Fig. 4-1, page 46). The specifications have to be adhered to as far as possible during installation, since the values influence the configuration of the system to some extent. Note A final (fine) adjustment of the components can only occur during startup (see Chapter 6.3, page 80). Prerequisite for this is the complete electrical installation of the system (Chapter 5, page 53). ICI890 Illumination with ICD890 Camera Distance of the illumination and the camera to the deflection mirror Reference point Deflection mirror Direction of transport Reading location 15° Distance of the deflection mirror to the reference point Abb. 4-6: Position of the components In order to facilitate component alignment, the deflection mirror and the unit consisting of the illumination and the camera are attached to the installation frame via 180° brackets. Bore holes for fastening screws Clamping screws Grooves for guide pins Pivotal point Bolts with slot nuts for attachment to the installation frame Abb. 4-7: 180° bracket for the deflection mirror and the illumination After loosening the clamping screws, the angle of the brackets can be adjusted from 0° to 180° as required. To achieve this, the clamping screws can be screwed into the various threaded holes. 8011325/0000/2006-10-24 © SICK AG · Division Auto Ident · Germany · All rights reserved 49