Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
the calibration is started via an inactive->active state change and stopped via an active>inactive state change. The active edge/state is specified via the IsDI1ForCalActiveLow
data point.
•
FALSE = Cal Edge Gated (default)
•
TRUE = Cal State Gated
For detailed digital proving configuration, refer to Section 4.1.5
Digital output
Select the function for which the digital output should be configured. Available options
are:
•
validity
•
flow direction
Figure 4-11: Digital outputs
The inverted operation option is useful if the output of the ultrasonic meter is reversed
from what a flow computer is expecting. This means that if the output normally outputs a
HIGH for a TRUE condition, selecting this checkbox will change the output to output a
LOW for a TRUE condition.