Download Prism Sound AD-2 Specifications
Transcript
Prism Sound dScope Series III Applications Manual Revision 1.11 Normalization attenuation. The test plots the output amplitude for the left channel every 500ms for 64s, measured through a 1kHz band-pass filter. Specification The sweep should show the output level dropping by 1dB every 2 seconds over a 30dB range. If compression is turned on, the output level should be flat for 3 seconds before beginning to drop. 5.2.1.5 Dynamic Range Compression dScope Configuration: Dynamic Range Compression Track 61.dsc Track Number: 61 Corresponding AP Test File: 5_dcomp Duration: 60 sec (each track) Test Signal Description: 400Hz Sweep, 0 to -60dBFS Channels: L, C, R, Ls, Rs Description This test verifies that dynamic range compression is functioning correctly during bass redirection. © 2003-2005 Prism Media Products Ltd 2.62