HeartView CT
CoronaryCARE
Indications:
This is a spiral scanning protocol, using a rotation time
of 0.5 s, ECG gating technique and a dedicated cardiac
filter which can reduce image noise thus provides the
possibility of dose reduction.
Topogram:
AP, 512 mm.
Approximately, from
the carina until the
apex of the heart.
A typical range of
12 cm covering the
entire heart can
be done in 28.7 s.
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Temporal resolution
Kernel
Increment
Image order
CTDIw
Effective dose
* Depends on heart rate.
24
CoronaryCARE
120
267
0.75 mm
1.0 mm
2.3 mm
0.5 sec.
up to 125 ms*
B30f
0.5 mm
cr-ca
22.4 mGy
Male: 3.8 mSv
Female: 5.5 mSv