Chapter 6. Maintenance
6.3 Cleaning the Mirror (cont.)
11. Visually inspect the mirror. Record any observation of gross contamination.
12. If the mirror does not appear to be clean, contact the factory for further assistance.
13. If the mirror appears to be clean, re-install it onto the measurement cell.
14. As you line up the mirror to the measurement cell, note the “key” for properly connecting the two pieces. The
mirror will align in only one position for mounting. Rotate the mirror assembly to line up the alignment pin with
the alignment pin slot (see Figure 81).
Alignment Pin Slot
Figure 81: Reinstalling the Mirror Assembly
15. Reinstall the six retaining bolts.
16. Tighten the six bolts in a star pattern. Hand tighten to achieve integral metal-to-metal contact.
17. Re-establish power to the Aurora H2O analyzer.
18. Re-establish the flow through the sample system.
19. If the Aurora H2O still indicates a Check Mirror, contact the factory for further assistance.
94
Aurora H2O User’s Manual