24 Measure Commands
:MEASure:UNITinterval
Command
This command is only available when the E2681A Jitter Analysis Software or the
N5400A/5401A Software is installed.
NOTE
:MEASure:UNITinterval <source>[,{AUTO | (SEMI,<data_rate>)}]
The :MEASure:UNITinterval command measures the unit interval value of
the selected source. Use the :MEASure:DATarate command/query to
measure the data rate of the source
<source>
{CHANnel<N> | FUNCtion<N> | WMEMory<N> | CLOCk | MTRend |
MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source.
<N>
CHANnel<N> is an integer, 1 - 4.
FUNCtion<N> and WMEMory<N> are:
An integer, 1 - 4, representing the selected function or waveform memory.
<data_rate>
Example
A real number representing the data rate.
This example measures the unit interval of channel 1.
myScope.WriteString ":MEASURE:UNITINTERVAL CHANNEL1"
Query
:MEASure:UNITinterval? <source>[,{AUTO | (SEMI,<data_rate>)}]
The :MEASure:UNITinterval? query returns the measured unit interval.
Returned Format
<value>
<result_state>
618
[:MEASure:UNITinterval] <value>[,<result_state>]<NL>
Unit interval of the source.
If SENDvalid is ON, the result state is returned with the measurement
result. See the :MEASure:RESults table in this chapter for a list of the
result states.
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
Related documents
Agilent Technologies Ininiium 9000 Technical data
Agilent Technologies Ininiium 90000 Specifications
Agilent Technologies Ininiium 90000 Specifications
Agilent Technologies 2000 X-Series Technical data
Agilent Technologies 86100A Technical data
Agilent Technologies InfiniiVision 3000 DSO-X 3054A Technical data
Agilent Technologies Ininiium 90000 Specifications
VDO MS 2000X - Technical data
(Agilent) 83495A Optical/electrical clock recovery
Agilent Technologies InfiniiVision 5000 Series Technical data
NCPlot™ v2.21 User Manual
NCPlot™ v2.26 User Manual