USER'S GUIDE____________________________________________________________________ Using Serial Line PRES and XPRES Command XPRES should be used if the value is changed frequently. Its value is not retained at reset, and when set to 0, last value set with PRES is used instead. Use the serial line and do the following: PRES [aaaa.a] XPRES [aaaa.a] where aaaa.a = Absolute process pressure (hPa) Example: >pres Pressure >pres 1010 Pressure > Table 30 : 1013.00 hPa ? : 1010.00 hPa Multiplication Factors From mbar Pa N/m2 mmHg torr inHg mmH2O inH2O atm at bar psia 1) 1) psia = psi absolute. To: hPa 1 0.01 1.333224 33.86388 0.09806650 2.490889 1013.25 980.665 1000 68.94757 Example: 29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa PFIX Use the PFIX command to select either P inputted or P measured. - When PFIX is On, fixed PRES value is used - When PFIX is Off, measured PRES value is used 100 _________________________________________________________________ M210796EN-E