Download National Instruments 6024E Switch User Manual
Transcript
Chapter 5 Calibration Other Considerations The CalDACs adjust the gain error of each analog output channel by adjusting the value of the reference voltage supplied to that channel. This calibration mechanism is designed to work only with the internal 10 V reference. Thus, in general, it is not possible to calibrate the analog output gain error when using an external reference. In this case, it is advisable to account for the nominal gain error of the analog output channel either in software or with external hardware. See Appendix A, Specifications, for analog output gain error information. © National Instruments Corporation 5-3 6023E/6024E/6025E User Manual