Table 4-1. HP E1429A/B Command Quick Reference (Cont’d).
Subsystem
Commands
Description
OUTPut:ECLTrg<n>:FEED <source >
Specifies the source of the synchronization
pulse routed to ECLTRG0 or ECLTRG1.
OUTPut:ECLTrg<n>[:STATe] <mode >
Enables/disables the routing of the
synchronization pulse.
OUTPut:EXTernal[1]:FEED <source >
Specifies the source of the synchronization
pulse routed to the "Ext 1" BNC port.
OUTPut:EXTernal[1][:STATe] <mode>
Enables/disables the routing of the
synchronization pulse.
OUTPut:TTLTrg<n >:FEED <source >
Specifies the source of the synchronization
pulse routed to a TTLTRG trigger line.
OUTPut:TTLTrg<n >[:STATe] <mode >
Enables/disables the routing of the
synchronization pulse.
READ
READ[<chan >]?
Returns readings from the specified
channel.
SENSe
[SENSe[<chan >]]:FUNCtion <function >
Selects the channel’s input port that will be
used.
[SENSe[<chan >]]:ROSCillator:EXTernal:FREQuency
<frequency >
Indicates the frequency of the external
reference source.
[SENSe[<chan >]]:ROSCillator:SOURce <source >
Selects the reference frequency source.
[SENSe[<chan >]]:SWEep:OFFSet:POINts <count >
Specifies the number of pre-arm readings.
[SENSe[<chan >]]:SWEep:POINts <count >
Specifies the total number of (pre- and
post-arm) readings taken during each arm
cycle.
[SENSe[<chan >]]:VOLTage[:DC]:RANGe <range >
Sets the digitizer measurement range.
[SENSe[<chan >]]:VOLTage[:DC]:RESolution?
Queries the digitizer reading resolution.
STATus:OPC:INITiate <state >
Controls whether *OPC, *OPC?, and *WAI
will complete immediately, or wait for the
measurement or data transfer to complete.
STATus:OPERation|:QUEStionable:CONDition?
Reads the settings of the condition register.
STATus:OPERation|:QUEStionable:ENABle <unmask >
Specifies which bits in the event register
are included in the summary bit.
STATus:OPERation|:QUEStionable[:EVENt]?
Reads the settings of the event register.
STATus:OPERation|:QUEStionable:NTRansition
<unmask >
Sets the negative transition mask.
STATus:OPERation|:QUEStionable:PTRansition
<unmask >
Sets the positive transition mask.
STATus:PRESet
Initializes the enable registers and
transition masks, and sets
STATus:OPC:INITiate ON.
SYSTem:ERRor?
Reads the error codes and messages in
the digitizer error queue.
SYSTem:VERsion?
Returns the SCPI conversion to which the
digitizer complies.
OUTPut
STATus
SYSTem
312 Command Quick Reference