Download HP E1429A User's Manual
Transcript
Table 4-1. HP E1429A/B Command Quick Reference (Cont’d). Subsystem Commands Description OUTPut:ECLTrg<n>:FEED <source > Specifies the source of the synchronization pulse routed to ECLTRG0 or ECLTRG1. OUTPut:ECLTrg<n>[:STATe] <mode > Enables/disables the routing of the synchronization pulse. OUTPut:EXTernal[1]:FEED <source > Specifies the source of the synchronization pulse routed to the "Ext 1" BNC port. OUTPut:EXTernal[1][:STATe] <mode> Enables/disables the routing of the synchronization pulse. OUTPut:TTLTrg<n >:FEED <source > Specifies the source of the synchronization pulse routed to a TTLTRG trigger line. OUTPut:TTLTrg<n >[:STATe] <mode > Enables/disables the routing of the synchronization pulse. READ READ[<chan >]? Returns readings from the specified channel. SENSe [SENSe[<chan >]]:FUNCtion <function > Selects the channel’s input port that will be used. [SENSe[<chan >]]:ROSCillator:EXTernal:FREQuency <frequency > Indicates the frequency of the external reference source. [SENSe[<chan >]]:ROSCillator:SOURce <source > Selects the reference frequency source. [SENSe[<chan >]]:SWEep:OFFSet:POINts <count > Specifies the number of pre-arm readings. [SENSe[<chan >]]:SWEep:POINts <count > Specifies the total number of (pre- and post-arm) readings taken during each arm cycle. [SENSe[<chan >]]:VOLTage[:DC]:RANGe <range > Sets the digitizer measurement range. [SENSe[<chan >]]:VOLTage[:DC]:RESolution? Queries the digitizer reading resolution. STATus:OPC:INITiate <state > Controls whether *OPC, *OPC?, and *WAI will complete immediately, or wait for the measurement or data transfer to complete. STATus:OPERation|:QUEStionable:CONDition? Reads the settings of the condition register. STATus:OPERation|:QUEStionable:ENABle <unmask > Specifies which bits in the event register are included in the summary bit. STATus:OPERation|:QUEStionable[:EVENt]? Reads the settings of the event register. STATus:OPERation|:QUEStionable:NTRansition <unmask > Sets the negative transition mask. STATus:OPERation|:QUEStionable:PTRansition <unmask > Sets the positive transition mask. STATus:PRESet Initializes the enable registers and transition masks, and sets STATus:OPC:INITiate ON. SYSTem:ERRor? Reads the error codes and messages in the digitizer error queue. SYSTem:VERsion? Returns the SCPI conversion to which the digitizer complies. OUTPut STATus SYSTem 312 Command Quick Reference Chapter 4