Download XPMPro 2.0 User's Manual - Organic Semiconductor Group Dr. Trixler

Transcript
XPMPro User's Manual
Data Analysis and Image Processing
the area enclosed by the box also encloses a few atomic sites and as the box is shifted over the
image, when it moves one atomic spacing, the correlation will increase because of the better
overlap of the atomic sites in the image and the atomic sites within the kernel. Also, the
correlation is very low at the step edges because the overlap is poor since the adsorbate is on a
terrace and the pixels at a step edge show large variations in z. This is responsible for the dark
regions running along each step edge in the image. To view the correlation map without the
crosses, click the Clear button.
Also, all the regions with a correlation above a threshold (set in the correlation window) are
averaged together to produce an optimized kernel (Figure 92). Since all of the features that
matched the defined kernel are averaged together signal to noise ratio is greatly reduced and a
more idealized or averaged appearance of the interesting feature can be seen. To manually
remove one of the locations from the averaging
calculation, make sure the Remove Points radio
button is checked and click on the cross to
remove that location.
A cross distribution of any two images of the
same size can be calculated using the 2D
Histogram menu item. When chosen, the cursor
changes shape and two images should be
selected. The Z values of the first data page are
plotted along the X axis, and the Z values of the
second image are plotted along the Y axis. The Z
values of the result correspond to the number of
pixels having the corresponding pair of Z values
in the original images. All the points in the 2D
correlation of an image with itself will lie on the
diagonal, indicating perfect correlation. This tool
is most useful for correlating contrast between
multiple images of the same area, such as
friction and topography, or tunnel current and
Figure 92
conductance, or left and right scans.
Two images can be subtracted to look for differences between them using the Image Scaled
Sum selection. After choosing the two images, a window will appear where the relative scaling
between the two images, c, is entered. Then each pixel of the first image is added to the same
pixel of the second image using the formula ZimageA + c*ZimageB The two images must have the
same units (both topography, both current, etc.) as well as identical pixel resolution (you cannot
subtract a 256x256 image from a 512x512 image). This can be very useful to notice the change
in the surface as a function of time or to subtract the left and right scans from each other when
acquiring friction data to visualize the different cantilever response in going from region 1 to
region 2 instead of region 2 to region 1. To subtract two images, use a value of (-1) for the
coefficient.
Image Processing
When any image processing procedure is selected, the type of window that appears depends on
the process that is chosen. For a calculation that has no user selectable parameters, the dialog
box will describe the calculation and contain two buttons. The first one will Cancel the
calculation and leave the original data intact. The second one will Calculate the result and then
RHK Technology
100