LabChip XT/XTe Instrument Description 215
Optics
The LabChip XT and XTe optics provide fluorescence detection
(532nm green excitation and 547-750nm green to red emission) for
DNA chips and methods.
Green Laser
The LabChip XT and XTe instruments include high intensity 532nm
green lasers to excite fluorescence in microfluidic chips.
Focusing, Alignment, and Calibration
The LabChip XT and XTe instruments are optically focused,
aligned, and calibrated at the factory. The optics does not require
alignment or focusing. An optical test chip is available for
instrument performance verification by the customer.
Photodiode Detector
The LabChip XT and XTe use a silicon photodiode detector for
fluorescence detection. Fluorescence signal from each channel is
acquired at a rate of 80Hz.
Optical Train
The fluorescence excitation and detection optical trains contain
lenses, fiber optics, and a high efficiency interference filter.
Barcode Reader
The LabChip XT and XTe instruments are equipped with an internal
Barcode Reader. The Barcode Reader reads the Caliper-applied
barcode on the edge of the chip.
The barcode reader is internal to the system and cannot be viewed
from the outside.
V2.1
LabChip XT/XTe User Manual
Caliper Life Sciences, Inc.