Bidirectional Transducer
The Micro Medical digital volume transducer consists of an acrylic tube with a vane
positioned between two swirl plates. The low inertia vane is attached to a stainless
steel pivot that is free to rotate on two jewelled bearings mounted at the centre of the
swirl plates. As air is passed through the transducer a vortex is created by the swirl
plates that causes the vane to rotate in a direction dependant upon the direction of
airflow. The number of rotations is proportional to the volume of air passed through
the transducer and the frequency of rotation is proportional to the flow rate. The
transducer housing consists of a main body that contains a pair of light emitting
diodes (LED’s) and phototransistors. The transducer is fixed to the mouthpiece
holder that pushes into the main body and is captured by an “O” ring seal. The
LED’s produce infrared beams, which are interrupted by the vane twice per
revolution. This interruption is sensed by the phototransistors. The output from the
collector of each phototransistor will be a square wave with a phase difference
between the two of + or - 90 degrees depending upon the direction of flow. The
square waves are detected by a microprocessor that measures the period of each
pulse and transmits that information to the main unit via a high-speed asynchronous
serial link.
There is no routine maintenance required for the transducer other than cleaning
according to the instructions in the operating manual.
Micro Medical Digital Volume Transducer
Rotating
vane
Infra red
emitter
Swirl
plate
Jewelled
bearing
Infra red
detector
Volume = k X No. of pulses
Volume proportional to the number of pulses
Flow proportional to the puse frequency
Flow = k / pulse period
4