4.0 MAINTENANCE
4.1 DAILY MAINTENANCE
At the end of each work session, it is advisable to carry out the following cleaning operations:
 Wash the serum plate under running water and dry with absorbent paper
 Empty the electrophoretic chamber and recycle the buffer; remove the floats from their housings,
rinse them with deionized water and leave them to dry.
 Remove the acetate supports from the strip racks.
 Clear the depositor by a soft bristle brush.
4.2 ROUTINE MAINTENANCE
It is advisable to wash the following parts more carefully at regular intervals:
 Migration chamber: fill the chamber, fill it with distilled water and leave it for an hour
 Clear the buffer side:
Remove the buffer tank and put a tank with a 10% hypochlorite solution. From the Reagents
Menu, click on the buffer 'Fill' button and after a couple of seconds press the 'Recycle' button.
 Immerse the serum plate in 10% hypochlorite solution for an hour and rinse with distilled water.