Download ElePhor 24S USER MANUAL
Transcript
4.0 MAINTENANCE 4.1 DAILY MAINTENANCE At the end of each work session, it is advisable to carry out the following cleaning operations: Wash the serum plate under running water and dry with absorbent paper Empty the electrophoretic chamber and recycle the buffer; remove the floats from their housings, rinse them with deionized water and leave them to dry. Remove the acetate supports from the strip racks. Clear the depositor by a soft bristle brush. 4.2 ROUTINE MAINTENANCE It is advisable to wash the following parts more carefully at regular intervals: Migration chamber: fill the chamber, fill it with distilled water and leave it for an hour Clear the buffer side: Remove the buffer tank and put a tank with a 10% hypochlorite solution. From the Reagents Menu, click on the buffer 'Fill' button and after a couple of seconds press the 'Recycle' button. Immerse the serum plate in 10% hypochlorite solution for an hour and rinse with distilled water. Page 33/52