Download Bedienungsanleitung
Transcript
Kommunikation Fieldbus / Meßgerät Index Busparameter Standard Transducer Block (TB) Parameter Description Mettler-Specific Parameters – Temperature 29 TEMP_WIRE_IMPEDANCE Sets the wire impedance of the temp. sensor. Typically 0 unless the wire of the sensor gets too long. 30 TEMP_SENSOR_CAL Desired temperature reading, used for temperature measurement calibration. The temp. value entered for a single point temp. standardization. Mettler-Specific Parameters – Impedance 31 GLASS_IMPEDANCE Sensor glass electrode impedance Value Status 32 REFERENCE_IMPEDANCE Sensor reference electrode impedance Value Status Mettler-Specific Parameters – Calibration 33 CAL_MAN_PH_POINT_1 pH of buffer solution used in a manual buffer calibration. 34 CAL_MAN_PH_POINT_2 pH of buffer solution used in a manual buffer calibration. 35 CAL_SLOPE_PH The slope of the pH electrode in % 36 CAL_ZERO_PH The zero offset resulting from a buffer calibration or a standardization 98 pH Transmitter 2100e FF