Cotton Swab
Handpiece
ENGLISH FRENCH
7-5 Cleaning (Rear of the handpiece) (X450STL)
Wipe all internal surfaces at rear of the handpiece with an alcohol-saturated cotton swab.
Fig. 9
7-6 Lubrication
NSK PANA SPRAY Plus
X450L/X450/X450KL/X450SL/X450BLED/X450WLED/X450WHL/X450MWL/X450STL
Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle onto the Pana Spray can.
3) Insert the Spray Nozzle in the rear of the handpiece. Hold the handpiece and spray for approximately
2-3 seconds. Continue until it runs clear from the handpiece head for at least 2 seconds. (Fig. 10)
Spray Nozzle
Handpiece
PANA SPRAY Plus
Fig. 10