Section 3: Performance Verification 6. Press the following softkeys; SETUP, NEXT, NEXT, and ANALOG. Press the 0-VOLT softkey. 7. Verify that the output voltage is +0.0 ± 0.025 VDC. 8. Leave the negative lead connected to pin 10 and verify 0.0 ± 0.025 VDC on pins 13 and 14. Note: 9. 3.3.3.7 Disconnect the voltmeter from the instrument. Operation on Battery Power 1. With the instrument operating on AC, turn on the backlight. 2. Disconnect the instrument from AC and verify that the AC/Battery Charging indicator turns off. 3. Verify that the instrument continues monitoring normally and that the low battery indicator is not lit. Note: 4. 3.3.4 If step 8 takes more than 2 minutes to complete, the analog output will time out. Repeat step 2 to initiate the analog output. If the low battery indicator is illuminated, perform the procedure outlined in step 3.3.1. Connect the instrument to AC and verify that the AC/Battery Charging indicator turns on and that the instrument is monitoring normally. General Operation The following tests are an overall performance check of the system: 3.3.4.1 • 3.3.4.1 LED Excitation Test • 3.3.4.2 Operation with a Live Subject LED Excitation Test This procedure uses normal system components to test circuit operation. A Nellcor Oxisensor ® II oxygen transducer, model D-25, is used to examine LED intensity control. The red LED is used to verify intensity modulation caused by the LED intensity control circuit. 1. Connect the monitor to an AC power source. 2. Connect an SCP-10 or MC-10 sensor input cable to the monitor. 3. Connect a D-25 sensor to the sensor-input cable. 4. Press the POWER ON/OFF button to turn the monitor on. 5. Leave the sensor open with the LEDs and photodetector visible. 6. After the monitor completes its normal power-up sequence, verify that the sensor LED is brightly lit. 7. Slowly move the sensor LED in proximity to the photodetector element of the sensor. Verify as the LED approaches the optical sensor, that the LED intensity decreases. 8. Open the sensor and notice that the LED intensity increases. 3-9