Download SERVICE MANUAL FOR MODEL 902 AUTOMATIC ANALYZER
Transcript
2.4.6 Check and Set Alarm of Each Data (1) Calibration Priority 1 2 3 4 5 6 Data Kind Potential of internal standard solution Low potential of standard solution High potential of standard solution Calibrator potential Slope value (for display) Concentration of internal standard solution Calibrator concentration Correction factor (2) Na K Cl Na K Cl Na K Cl Na K Cl Na K Cl Na K Cl Noise error Level error Na K Cl Na K Cl Calculation disabled ADC abnormal Sample short Noise error ↑ ↑ ↑ ↑ ↑ ↑ Slope abnormal Preparation abnormal Internal standard concentration abnormal Calibration abnormal Calculation disabled Level error Calibration Calculation abnormal disabled Calculatio n disabled Routine Sample Measurement (1st), Stat Sample Measurement (1st) and Control Sample Measurement Priority 1 2 3 4 5 6 Data Kind Sample concentration Na K Cl ADC abnormal Sample short Noise error Level error Sample value abnormal If multiple data alarms concur, the one with the highest priority will be indicated. 2 - 60 Calculatio n disabled