SECTION 5 MAINTENANCE AND SERVICE TESTS
5.2.2
AC POWER CORD, RETAINER, AND VELCRO STRAP INSPECTION
Inspect the power cord, power cord retainer, and Velcro strap at least once every
12 months.
To inspect the power cord and retainer, see Figure 5-2, and proceed as follows:
1. Turn off the infuser and disconnect the device from AC power.
2. Remove the retainer and power cord as described in Section 7.2.5.
3. Inspect the retainer for cracks, breaks, or missing parts of the retainer body.
If any damage is found, replace the retainer.
4. Inspect both ends of the power cord for any signs of electrical arcing, burn marks,
or heat scorching or melting. If any damage is observed, replace the power cord.
5. Inspect the plug end of the power cord for bent blades or a bent or missing ground
pin. If any damage is observed, replace the power cord.
6. Inspect the Velcro strap for damage. If any damage is observed, replace the strap.
7. When inspections are completed, reassemble the power cord to the infuser
as described in Section 7.2.5.
8. Connect the infuser to AC power and confirm the AC indicator is lit.