Troubleshooting
When the short residual resistance is out of the limit, check the xture connection, the
shorting plate, and the surface of the electrodes. If the surface of the electrode is damaged,
electrode assembly replacement is required for restoring proper electrical performance.
When the short residual inductance is out of the limit, check the angle of the semi-rigid cables
in back of the electrode assemblies. The angle should be close to 90 . Also check the shield of
the semi-rigid cables are connected only at the copper round plate.
Caution
4-6
Service
Do not polish the surface with powder to protect the electrode.