Download CardioSoft - Davis Medical Electronics

Transcript
Troubleshooting Tips: Frequently Asked Questions
Question: Access to DEMO CardioSoft Web?
Answer:
http://194.45.46.68
Login with webuser1 and password webuser1.
Problem:
If in exercise tests with the EC1200 the stage time for the
PRETEST phase is set to 0 minutes, the error message “No
communication with ergometer/treadmill” appears in about
every second exercise test after initiation of the modality
and before the patient start pedalling. This problem only
occurs with CardioSoft V4.2 (not with V4.14).
Solution:
Change the PRETEST phase of the protocol from 0 minutes to 99 minutes and select
“Phase >>” when the patient start pedalling. With this method, the connection will
not be interrupted.
Problem: Patient assignment problems of CardioSoft in
conjunction with MASTERSCOPE V4.52?
Solution:
Mr. Eirich knew about the patient assignment problem of CardioSoft V4.14 and
Masterscope V 4.52.
The problem has been resolved with CardioSoft V4.2.
Mr. Eirich only released CardioSoft V4.2 in conjunction with V4.52 and higher.
The functionality of transferring patient data from the medical reimbursement program to the JAEGER Masterscope program via CardioSoft is not implemented in
V4.14 or V4.2, this functionality is implemended in CardioSoft V5.x.
Network installation using cardiouser as the domain user
does not work.
What can be done ?
For Cardiosys to run in the Windows2000 environment, the registered user must be
a member of the local group of main users. This applies to both local users and
domain users. The user must have the right to access the Cardio installation directory and the Temp directory. We assume you have verified the latter requirement. So
the user rights are probably the root cause of the problem.
Log on to the Cardiosys client as the Administrator (local), add the user
“?Domäne?\cardiouser” to the local user group of main users. Repeat this configuration at each CardioSys client.
Revision C
CardioSoft
2014708-001