Transcript
Troubleshooting Performance test failure troubleshooting Performance test failure troubleshooting This section describes the adjustment and troubleshooting procedures used when the E4980A/ E4980AL fails the performance tests. If the performance of the instrument is critical for the test limits and seems to be adjustable, perform first the adjustment(s) related to the failed test. When the test result is far from the tolerance of the test or the performance is not adjustable, isolate the faulty assembly in accordance with the "Performance tests failure troubleshooting procedure". Recommended adjustment for performance test failure Table 4-13 shows the recommended adjustments when the performance test fails. Select the adjustment program corresponding to the recommended adjustment and perform the adjustment. Table 4-13 Test Num. Recommended adjustment for performance test failure First failed test Recommended adjustment A1 A2 A3 1 Frequency Accuracy Test 2 DC Source Level Accuracy Test 3 DC Bias Level Accuracy Test 4 DC Bias V Level Monitor Accuracy Test 5 DC Bias I Level Monitor Accuracy Test 6 AC Signal Level Monitor Accuracy Test 7 AC Signal Level Accuracy Test 8 Impedance Measurement Accuracy Test 9 DCR Measurement Accuracy Test 4. Troubleshooting Chapter 4 PPMC 63