Download - Frank`s Hospital Workshop
Transcript
BT-300 Service Manual V2.0 J6 S IG N A L 1 2 5 2 6 7 -2 Pin 1 2 Name Signal GND I/O I - Description Mark enable signal Ground D. Doppler 2 control signal (only BT-300 dual) D O P2_P VO L2_0 VO L2_1 D O P2_S D O P2_H R AVP5 2 4 6 8 2 4 6 8 J3 J5 AN AL_C O N A N A L _ S IG 1 3 5 7 1 3 5 7 1 3 5 7 S IG N A L 2 4 6 8 J4 TR AN S D U C ER VO L2_2 D O P2_Q D O P2_I Name Signal DOP2_P VOL2_0 VOL2_1 VOL2_2 DOP2_Q DOP2_I DOP2_S DOP2_HR AVP5 AVP12 AVP12 Description Transmitting and receiving ultrasound signal Digital pulse for transmitting Volume control signal 0 Volume control signal 1 Volume control signal 2 180° control signal for demodulation 0° control signal for demodulation Doppler sound signal Doppler heart rate signal +5V power +12V power E. Speaker Connector 19