Download Agilent 1200 Series Variable Wavelength Detector G1314D /G1314E
Transcript
13 Repair Changing the Type and Serial Number Recover Instructions Overview The following situations may come up where the instrument is no longer usable due to • an incorrect type (product number) entry after the replacement of a main board of the module. • load of wrong firmware based on the wrong type. NOTE With firmware A.06.02/B.01.02 and above a wrong type cannot be entered. The entry is checked against the board revision. The wrong type (product number) could be • incorrect, but a valid 1100/1200 series module number • incorrect and invalid 1100/1200 series module number (any name) Based on above, the User Interfaces react differently. Table 25 Recover From Wrong Type User Interface incorrect but valid type incorrect but valid type incorrect and invalid type Example Conditions correct type = G1315B entered type = G1314B correct type = G1315B entered type = G1314B plus wrong firmware from G1314B correct type = G1315B entered type = G1319B ChemStation “Recover with Agilent Diagnostic Software (Type & Firmware)” on page 226 shows the incorrect product number Interface shows the settings of the G1314B Type can be changed via command line as described under “Recover with Agilent ChemStation” on page 227 does not show the module NO access to the module is possible Use “Recover with Agilent Diagnostic Software (Type & Firmware)” on page 226 does not show the incorrect product number NO access at all to the module is possible Use “Recover with Agilent Diagnostic Software (Type & Firmware)” on page 226 224 1200 Series VWD Service Manual