Section 6: Mobile Unit Transport Procedure
The medical system requires the HVAC system to be supplied power at all
times. During transit of the mobile unit via the on board generator and when
the unit is in the parked position via shore power.
The landing/stabilizing legs are not to be used to raise the mobile unit off the
ground. The legs are meant to level the vehicle only. If the legs are used in an
attempt to raise the unit off the ground, serious damage may occur.
A checklist can be found in Appendix A that may be used as a guideline for
the following procedure.
6.1
Secure Medical Equipment per OEM Requirements
All medical equipment must be secured prior to the transport of the mobile unit. Refer to the OEM
instructions for securing all medical equipment.
Catheterization Lab
The main medical components to be secured in the medical procedure room are the C
arm, table, operating light, radiation shield, and monitor suspension.
1. Position the C arm and clamp the transport pin in place as shown in Figure 28-A.
2. Pan the table forward and down and clamp in place with the table transport rod as
shown in Figure 28-B.
3. Rotate the Radiation Shield and Monitor suspension arms to the rear and orient
them as shown in Figure 28-C and Figure 28-D.
4. Using the foam cushion and straps provided, secure the Radiation Shield and
Monitors in place on the table as shown in Figure 28-E and Figure 28-F.
5. Using the foam cushion and straps provided, secure the MedRad Injector to the
table as shown in Figure 28-G.