Download Brukerveiledning MOLIFT Smart

Transcript
Replacing the battery
• Open the chassis.
Observe how the cables are placed inside the
lifter before disconnecting.
• Disconnect the battery cable connector (red
and black cable) from main circuit board.
• Lift the main circuit board enouch to
disconnect the cable for temperature sensor.
• Pull the battery straight out.
• Assembly is done in reverse order.
Make sure that no cables are squeezed
between the battery and the ribs/walls of the
chassis.
• Fix the cables with cable strips.
Replacing of the main circuit board
The circuit board can be changed when all
settings and data for this lifter is copied from
the old circuit board to the new. If there is any
trouble when copying the settings, please
send the lifter to Molift Group AS or your
dealer.
Observe how the cables are placed inside the
lifter before disconnecting.
• Use the Molift Servicetool , connect to lifter
and copy from the lifter to the PC first. Then
connect to the new main circuit board and
copy on request.
• Open the chassis and remove the battery.
• Pull the circuit board carefully out and loosen
the six cable connectors.
• When mounting a new main circuit board,
connect the cables in the following order:
1 Emergency lowering cable
2 Temperature sensor cable
3 Cable to brake (motor 2)
4 Cable for switch up/down buttons.
Push the diodes and fuses on the new board
carefully aside to prevent it from colliding
with ribs/walls on the chassis.
Microswitches
Up/down
buttons
Brake
Emergency
lowering
Temp.sensor
Motor 1 (Liftingmotor)
Page 9 of 24