VALUEGEN ® TECHNICAL MANUAL 35 mA Station Selection You have a maximum of 11 mA Stations which you can program on the V-50/V-650 generator, 10 on the V-40/V-550, 9 on the V-30/V-425, 8 on the V-20/V325 and 7 on the V-10. You may use any combination between the large and the small filaments. NOTE: In the Time Display there are three (3) numbers which will be displayed. 0 1 2 Not Programmed or Not Available Progammed on Small Filament Programmed on Large Filament Small mA Station Selection Review the factory settings in Table 4-1. If the settings are ok no further action is required. If you wish to make a change in any of the selections use the following procedure. With the Setup Location for the mA station showing in the mAs display, select the desired programming in the mA display. Your options are: 0 1 2 Not Programmed or Not Available Progammed on Small Filament Programmed on Large Filament Large mA Station Selection Review the factory settings in Table 4-1. If the settings are ok no further action is required. If you wish to make a change in any of the selections use the following procedure. With the Setup Location for the mA station showing in the mAs display, select the desired programming in the mA display. Your options are: 0 1 2 Not Programmed or Not Available Progammed on Small Filament Programmed on Large Filament Filament Standby Number The generator has the capability of adjusting the standby current through the large and small filaments independently of each other. The filament standby current is adjusted by increasing or decreasing the number displayed on the Time Display of the Operator’s Console during Set Up. This number does not directly correlate to an actual amount of filament current. It is subjective and the adjustment procedure is augmented by visually inspecting each filament and monitoring the FIL I test point on the BBU pcb as you make the adjustment. FIL I Test Point Figure 4-7 FIL I Test Point Copyright 2007 GTR LABS, Inc. GND