VALUEGEN ® TECHNICAL MANUAL
35
mA Station Selection
You have a maximum of 11 mA Stations which you can program on the V-50/V-650 generator, 10 on the V-40/V-550,
9 on the V-30/V-425, 8 on the V-20/V325 and 7 on the V-10.
You may use any combination between the large and the small filaments.

NOTE: In the Time Display there are three (3) numbers which will
be displayed.
0
1
2
Not Programmed or Not Available
Progammed on Small Filament
Programmed on Large Filament
Small mA Station Selection
Review the factory settings in Table 4-1. If the settings are ok no further action is required.
If you wish to make a change in any of the selections use the following procedure.
With the Setup Location for the mA station showing in the mAs display, select the desired programming in the mA
display.
Your options are:
0
1
2
Not Programmed or Not Available
Progammed on Small Filament
Programmed on Large Filament
Large mA Station Selection
Review the factory settings in Table 4-1. If the settings are ok no further action is required.
If you wish to make a change in any of the selections use the following procedure.
With the Setup Location for the mA station showing in the mAs display, select the desired programming in the mA
display.
Your options are:
0
1
2
Not Programmed or Not Available
Progammed on Small Filament
Programmed on Large Filament
Filament Standby Number
The generator has the capability of adjusting the standby current through the large and small filaments independently
of each other.
The filament standby current is adjusted by increasing or decreasing the number displayed on the Time Display of the
Operator’s Console during Set Up.
This number does not directly correlate to an actual amount of filament current. It is subjective and the adjustment
procedure is augmented by visually inspecting each filament and monitoring the FIL I test point on the BBU pcb as
you make the adjustment.
FIL I Test Point
Figure 4-7 FIL I Test Point
Copyright 2007 GTR LABS, Inc.
GND