SECTION 1
Performance Verification
1.5.8
External
DC Bias Voltage
1. Connect the LCZ meter to the DC power supply, capacitor, and DMM, as shown in Figure 1-6.
2. Select the DCV function and auto-ranging
on the
DMM.
3. Set LCZ meter operating modes as follows:
m DISPLAY: C
LEVEL: 50mVrms
BIAS INT/EXT: EXT (rear panel switch)
BIAS: ON or OFF as indicated below.
4. For the Model 3322 only, set the unit to RANGE 2 as
follows:
A. Measure the value of the 10&F capacitor in the
RANGE AUTO mode (DC BIAS OFF).
B. Verify that the Model 3322 is properly measuring the capacitor.
C. Press the RANGE/AUTO
key, and check that
the AUTO LED goes off. The unit is now set to
RANGE 2.
‘igure l-6.
5. Check the voltage across the capacitor with BIAS
OFF, and with BIAS ON and a 0.35V external bias
voltage setting on the DC power supply.
6. Disconnect the DMM, and measure the capacitor
value. Verify that the reading is stable to within 12
digits of the center value (4 digits of span).
7. Connect the DMM to measure the voltage across the
capacitor.
Range
WARNING
Hazardousvoltage
k-3OV) will be used in the
following
steps. Do not touch the capacitor
until the test is complete and the capacitor is
discharged.
8. Gradually increase the external DC power supply to
+4ov.
9. Verify that the external bias fuse does not blow with
40V applied.
10. Gradually decrease the external DC power supply to
OV, then disconnect the capacitor.
Connectionsfor External Bias Voltage RangeTests
1-7