Spirometry (Gas Module SE with Spirometry Only) Overview Theory Of Operation FIGURE 1-24 PVX measuring unit The measurement is based on measuring the kinetic gas pressure and is performed using the Pitot effect. A pressure transducer is used to measure the Pitot pressure. The signal is then linearized and corrected according to the density of the gas. Speed of the flow is calculated from the pressure and Vt (Tidal Volume) is integrated from it. The PVX unit consists of airway connections, two pressure transducers, valves and preamplifiers. The preamplifiers are connected to the A/D-converter on the main CPU module. A patient’s breathing flow passing through the adapter creates a pressure difference. This pressure difference is measured by pressure transducer, B1. Overpressure and negative pressure in airways are measured by another pressure transducer, B2. 1 - 26 0070-10-0522 Gas Module Service Manual Addendum