Download Service Manual - Agilent Technologies
Transcript
Repair Fault Troubleshooting No Serial Port activity. Cause: Processor board faulty. No Tx signal with PRBS or fixed pattern. Cause: Processor board, or North connector assembly or cable faulty. Or Rx only Test Mode selected. No TX signal with VF tone pattern. Cause: PSU board, Processor board, or North connector assembly or cable. Action: 1.Change the TX pattern setting to VF tone. If no Tx signal, suspect the PSU board. No speaker output. Cause: PSU board, Processor board, or Keymat faulty. Action: 1.Change the TX pattern setting to VF tone. If no speaker output, change the PSU board. If there is still no speaker output, suspect the Keymat. 2. Change the TX pattern setting to PRBS, select Beep on Error and generate a 1x10-3 errors. If there is still no speaker output, suspect the Processor board. Instrument switches off soon after reaching the Tx/Rx page. Cause: Flat battery. Action: Apply external power for 24 hours. Instrument switches off soon after reaching the Tx/Rx page with a “charged” battery. Cause: Battery or PSU board faulty. Action: Apply external power and listen for “buzz” sound from the instrument in quiet conditions. If there is no “buzz” sound, suspect the PSU board. 3-5