If the readings fail the specied limits, perform the \A14 VCO Exercise."
A14 VCO Range Check with Oscilloscope
1. Remove the W9 HI OUT cable (A14J1 to A7) from the A7 assembly and connect it to an
oscilloscope set for 50 ohm input impedance. Switch on the analyzer.
2. Press 4PRESET5 4SYSTEM5 SERVICE MENU SERVICE MODES FRACN TUNE ON to activate the
FRACN TUNE service mode. See Chapter 10, \Service Key Menus and Error Messages,"
for more information on the FRACN TUNE mode.
3. Vary the fractional-N VCO frequency with the front panel knob and check the signal with
the oscilloscope. The waveform should resemble Figure 7-17, Figure 7-18, and Figure 7-19.
If the fractional-N output signals are correct, continue source troubleshooting by skipping
ahead to \A7 Pulse Generator Check."
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Figure HO107 here.
Figure 7-17. 10 MHz HI OUT Waveform from A14J1
DRAFT
3/21/106 15:13
Source Troubleshooting 7-23