Laboratory Facilities
The nitrous oxide supply must be regulated to maintain the
instrument operating pressure over the full range of normal flow
rates. A dual-stage regulator should be used. This is most important
for instruments with programmable flow rates because a slow
response from the regulator may cause the instrument to shut down
during a demand for increased flow. A shutdown of this kind is safe
and does not imply a fault in the instrument, but will interrupt your
work.
Nitrous oxide is stored as a liquid. Evaporation across the regulator
may cause it to freeze at high flow rates. This can cause erratic flame
conditions and in extreme cases, may cause a flashback. You can
purchase a pre-heater, which will prevent freezing. Your Agilent sales
and service office can advise you and can supply these devices:
Table 12. Nitrous oxide supplies
Region
Description
North America
Regulator with integral thermostatted heater
Continental Europe
Heater, for connection between cylinder and regulator
Figure 18. Nitrous oxide supply connection adapters
62
Agilent AA Spectrometer Systems Site Preparation Guide